Memory full!

Delete:
Escape:

DELETE
SAMPLE

Cause

Lamp is defect.

Remedy

Replace the lamp.

20° Lamp defect !

Continue:

OPERATE

Please Recalibrate!

Continue:
Calibration:

SET UP
OPERATE

Cause

Lamp aging or loss of stored calibration
values.

Remedy

Re-calibrate.

Cause

The storage capacity is exhausted
(600 memory locations).

Remedy

If possible, transfer stored data to a PC.

Memory full!

Delete:
Escape:

DELETE
SAMPLE

Cause

Lamp is defect.

Remedy

Replace the lamp.

20° Lamp defect !

Continue:

OPERATE

Please Recalibrate!

Continue:
Calibration:

SET UP
OPERATE

Cause

Lamp aging or loss of stored calibration
values.

Remedy

Re-calibrate.

Cause

The storage capacity is exhausted
(600 memory locations).

Remedy

If possible, transfer stored data to a PC.