GLOSS CALIBRATION

Measure High-Gloss

--> OPERATE

Escape:

SAMPLE

GLOSS CALIBRATION

20° Calibration ok

90.2

60° Calibration ok

93.3

85° Calibration ok

99.1

set up

operate

Move down the sample support table
using the hand lever and insert black
high-gloss standard.
If your

haze-gloss is equipped with a

motorized sample table you can
alternatively use the footswitch.

The calibration cycle for the single
angle geometries is now performed
automatically.

SET UP MENU
Haze Compens.:
Calibration
Measure autom:
Target:
Lower:
Decimal Point:
Escape:

on

on
on
4cm
on
SAMPLE

GLOSS CALIBRATION

20° Calibration ok

90.2

60° Calibration ok

93.3

85° Calibration ok

99.1

set up

operate

Move down the sample support table
using the hand lever and insert black
high-gloss standard.
If your

haze-gloss is equipped with a

motorized sample table you can
alternatively use the footswitch.

The calibration cycle for the single
angle geometries is now performed
automatically.

SET UP MENU
Haze Compens.:
Calibration
Measure autom:
Target:
Lower:
Decimal Point:
Escape:

on

on
on
4cm
on
SAMPLE

GLOSS CALIBRATION

Measure High-Gloss

--> OPERATE

Escape:

SAMPLE

• Select Calibration
Routine

• Call Calibration Routine

Note
Make sure that the glass standard is
clean and has no scratches!

• Select Calibration
Routine

• Call Calibration Routine

Note
Make sure that the glass standard is
clean and has no scratches!