ALTANA Haze-Gloss Manual User Manual
Page 14

GLOSS CALIBRATION
Measure High-Gloss
--> OPERATE
Escape:
SAMPLE
GLOSS CALIBRATION
20° Calibration ok
90.2
60° Calibration ok
93.3
85° Calibration ok
99.1
set up
operate
Move down the sample support table
using the hand lever and insert black
high-gloss standard.
If your
haze-gloss is equipped with a
motorized sample table you can
alternatively use the footswitch.
The calibration cycle for the single
angle geometries is now performed
automatically.
SET UP MENU
Haze Compens.:
Calibration
Measure autom:
Target:
Lower:
Decimal Point:
Escape:
on
on
on
4cm
on
SAMPLE
GLOSS CALIBRATION
20° Calibration ok
90.2
60° Calibration ok
93.3
85° Calibration ok
99.1
set up
operate
Move down the sample support table
using the hand lever and insert black
high-gloss standard.
If your
haze-gloss is equipped with a
motorized sample table you can
alternatively use the footswitch.
The calibration cycle for the single
angle geometries is now performed
automatically.
SET UP MENU
Haze Compens.:
Calibration
Measure autom:
Target:
Lower:
Decimal Point:
Escape:
on
on
on
4cm
on
SAMPLE
GLOSS CALIBRATION
Measure High-Gloss
--> OPERATE
Escape:
SAMPLE
• Select Calibration
Routine
• Call Calibration Routine
Note
Make sure that the glass standard is
clean and has no scratches!
• Select Calibration
Routine
• Call Calibration Routine
Note
Make sure that the glass standard is
clean and has no scratches!