• Set reflectometer value:

• Press the

set up key to

change the number above
the arrow, cycling from

0 - 9.

• Confirm the selected
number using the

operate key. The arrow

jumps to the next number.

• Measure mirror reflection
standard.

haze-gloss calibration is

performed against your

own standard.

CUSTOM SPEC. CALIB.
Calib. Value 20°

* * * 1 2 3 0 * * *

Change Value:
Confirm:
Escape:

SETUP
OPERATE
SAMPLE

CUSTOM SPEC. CALIB.

Measure Custom Tile

20° : 1000


--> OPERATE

Escape:

SAMPLE

set up

operate

operate

Note

If you define very high standard values,
limit errors might occur during the
measurement as the numerical range is
limited to 500 gloss units. In this case,
4 stars will be displayed instead of the
reflectometer value.

• Set reflectometer value:

• Press the

set up key to

change the number above
the arrow, cycling from

0 - 9.

• Confirm the selected
number using the

operate key. The arrow

jumps to the next number.

• Measure mirror reflection
standard.

haze-gloss calibration is

performed against your

own standard.

CUSTOM SPEC. CALIB.
Calib. Value 20°

* * * 1 2 3 0 * * *

Change Value:
Confirm:
Escape:

SETUP
OPERATE
SAMPLE

CUSTOM SPEC. CALIB.

Measure Custom Tile

20° : 1000


--> OPERATE

Escape:

SAMPLE

set up

operate

operate

Note

If you define very high standard values,
limit errors might occur during the
measurement as the numerical range is
limited to 500 gloss units. In this case,
4 stars will be displayed instead of the
reflectometer value.