ALTANA Haze-Gloss Manual User Manual
Page 16

• Set reflectometer value:
• Press the
set up key to
change the number above
the arrow, cycling from
0 - 9.
• Confirm the selected
number using the
operate key. The arrow
jumps to the next number.
• Measure mirror reflection
standard.
•
haze-gloss calibration is
performed against your
own standard.
CUSTOM SPEC. CALIB.
Calib. Value 20°
* * * 1 2 3 0 * * *
Change Value:
Confirm:
Escape:
SETUP
OPERATE
SAMPLE
CUSTOM SPEC. CALIB.
Measure Custom Tile
20° : 1000
--> OPERATE
Escape:
SAMPLE
set up
operate
operate
Note
If you define very high standard values,
limit errors might occur during the
measurement as the numerical range is
limited to 500 gloss units. In this case,
4 stars will be displayed instead of the
reflectometer value.
• Set reflectometer value:
• Press the
set up key to
change the number above
the arrow, cycling from
0 - 9.
• Confirm the selected
number using the
operate key. The arrow
jumps to the next number.
• Measure mirror reflection
standard.
•
haze-gloss calibration is
performed against your
own standard.
CUSTOM SPEC. CALIB.
Calib. Value 20°
* * * 1 2 3 0 * * *
Change Value:
Confirm:
Escape:
SETUP
OPERATE
SAMPLE
CUSTOM SPEC. CALIB.
Measure Custom Tile
20° : 1000
--> OPERATE
Escape:
SAMPLE
set up
operate
operate
Note
If you define very high standard values,
limit errors might occur during the
measurement as the numerical range is
limited to 500 gloss units. In this case,
4 stars will be displayed instead of the
reflectometer value.