• Call SET UP Menu

SET UP OPERATION

Cursor: SET UP
Change: OPERATE



Continue: SET UP

5. Calibration

• Call SET UP Menu

SET UP OPERATION

Cursor: SET UP
Change: OPERATE



Continue: SET UP

5. Calibration

Calibration of the

haze-gloss is seldom

necessary due to state of the art
technology and a reference beam
principle. We recommend a calibration
interval of

≥ 2 months.

Calibration is also recommended

• if the ambient temperature differs
extremely from the normal room
temperature.

• if "Please Recalibrate" appears on the
display. (The unit automatically de-
tects lamp aging by 4 %)

set up

Calibration Procedure

Calibration Procedure

set up

Calibration of the

haze-gloss is seldom

necessary due to state of the art
technology and a reference beam
principle. We recommend a calibration
interval of

≥ 2 months.

Calibration is also recommended

• if the ambient temperature differs
extremely from the normal room
temperature.

• if "Please Recalibrate" appears on the
display. (The unit automatically de-
tects lamp aging by 4 %)