Measuring Geometries:

20° & haze, 60° , 85°

Illuminated Area (mm/in.):

20° & haze

60°

85°

ø 15

15 x 27 8 x 60

ø 0.6

0.6 x 1.0 0.3 x 2.4

Measuring Range:
Gloss

0 - 2000 gloss units

Haze

0 - 2500 gloss units

Range change
to mirror reflection:

automatic or preselectable

Display Resolution:

0.1 unit (range: 0-100 units)

1 unit (range: > 100 units)

Accuracy**:
Gloss
Repeatability

± 0.05 units*

Reproducibility

± 0.3 units*

Absolute Accuracy ± 1 unit
Long-term Stability ± 0.2 units*

(> 10000

measurements)

Temperature Drift

0.02 units/°C*

Haze
Repeatability

± 0.3 units*

Reproducibility

± 2.0 units*

Long-term Stability ± 0.3 units*

(> 10000 measur.)

Temperature Drift:

0.2 units/°C*

Measuring Time:

1.5 - 4 sec.

(mode dependent)

Light Source:

tungsten lamp

(5 V / 190 mA)

Detector:

silicon photocell

V(Lambda) filtered

Display:

LCD

with back lighting

Statistics Memory: 9 x 600 meas.

battery backed

data protection for

10 years

Interface:

RS 232 C for

control and data

evaluation via PC

Connections:

RS 232 C

9-pin Sub-D

footswitch

15-pin Sub-D

Power Supply:

115/230 V

50/60 Hz

Power Requirement: 50 VA

Fuses:

1 AT (230 V)

1.6 AT (115 V)

11. Technical Data

11. Technical Data

Measuring Geometries:

20° & haze,

60° , 85°

Illuminated Area (mm/in.):

20° & haze 60°

85°

ø 15

15 x 27 8 x 60

ø 0.6

0.6 x 1.0 0.3 x 2.4

Measuring Range:

Gloss

0 - 5000 gloss units

Haze

0 - 2500 gloss units

Range change

to mirror reflection:

automatic or preselectable

Display Resolution:

0.1 unit (range: 0-100 units)

1 unit (range: > 100 units)

Accuracy**:

Gloss

Repeatability

± 0.05 units*

Reproducibility

± 0.3 units*

Absolute Accuracy ± 1 unit

Long-term Stability

± 0.2 units*

(> 10000

measurements)

Temperature Drift

0.02 units/°C*

Haze

Repeatability

± 0.3 units*

Reproducibility

± 2.0 units*

Long-term Stability

± 0.3 units*

(> 10000 measur.)

Temperature Drift: 0.2 units/°C*

Measuring Time:

1.5 - 4 sec.

(mode dependent)

Light Source:

tungsten lamp

(5 V / 190 mA)

Detector:

silicon photocell

V(Lambda) filtered

Display:

LCD

with back lighting

Statistics Memory: 9 x 600 meas.

battery backed

data protection for

10 years

Interface:

RS 232 C for

control and data

evaluation via PC

Connections:

RS 232 C

9-pin Sub-D

footswitch

15-pin Sub-D

Power Supply:

115/230 V

50/60 Hz

Power Requirement: 50 VA

Fuses:

1 AT (230 V)

1.6 AT (115 V)