ChemoMetec SP-100 User Manual
Page 70
11 Equipment and
Accessories
52
Cord type min. H05RR-F or min. H03VV-F or min. H03VVH2-F, rated min. 60C, 3x0.75
mm2. Provided with molded-on grounding-type attachment Plug, rated min. 250 Vac,
min. 2.5 A. Opposite end terminates in molded on, IEC320 style connector, rated min.
250 Vac, min. 2.5 A.
Detachable power supply cord set for the specified Class II power supply:
For US 125 Vac
UL listed, type SPT-2 or SVT, rated min. 60C, 18 AWG, 2 conductors. Provided with
molded-on ungrounding-type (NEMA 1-15P) attachment Plug, rated min. 125 Vac, min.
2.5 A. Opposite end terminates in molded on, IEC320 style connector, rated min. 125
Vac, min. 2.5 A.
For Europe 250Vac
Cord type min. H05RR-F or min. H03VV-F or min. H03VVH2-F, rated min. 60C,
2x0.75mm2. Provided with molded-on ungrounding-type attachment Plug, rated min.
250 Vac, min. 2.5 A. Opposite end terminates in molded on, IEC320 style connector,
rated min. 250 Vac, min. 2.5 A.
The Mains supply cord and plug of the external power supply shall comply with any
national regulations.
The user shall be made aware of that, if the NucleoCounter SP-100 and the external
power supply is used in a manner not specified by the manufacturer, the protection
provided by the NucleoCounter SP-100 and the external power supply may be impaired.
Contact ChemoMetec A/S for information on order of a specific Power supply cord listed
above.