6 Description of a typical Analysis Procedure (Total Count)
37
4.
Immediately, i.e. max. 10 seconds after the dispensing step, a portion of the
lyzate mixture shall be aspirated into a SP1-Cassette (see
Figure 65
). The tip of
the cassette should be immersed below the surface of the sample during
aspiration. Apply a consistent pressure to the piston and press the piston all the
way until it reaches the finger grips of the cassette.
Avoid touching the
window (clear area) of the measurement chamber....
Figure
Figure
Figure
Figure 65
65
65
65.
.
.
. Loading the cassette with lyzate mixture
5. Once the sample has been aspirated into the SP1-Cassette, open the lid and
insert the cassette in the insertion slit of the NucleoCounter SP-100. Close the lid
and press the “
Run” button on the instrument in order to initiate the analysis
(see
Figure 66
).
Figure
Figure
Figure
Figure 66
66
66
66.
.
.
. Insertion of cassette (A), closing the lid (B) and pressing the “
Run” button (C)
A
B
C