ChemoMetec SP-100 User Manual
Page 64

9 Technical specifications
46
analysis is presented in
Millions of Cells per ml
of the initial Sample. The result is
always shown on the NucleoCounter display. If connected to a PC with
NucleoView installed the result is also displayed on the PC monitor. The viability
in percent can also be displayed on the PC-monitor if both a Total Count and a
Non-Viable Count are performed.
Operation
Menu-controlled by means of keyboard and LCD display.
Analysis time
When pressing “Run
Run
Run
Run” on the NucleoCounter SP-100 the result will be displayed
within:
30 seconds for a Total Count
60 seconds for a Non-Viable Count
Physical data
Weight
Height
Width
Depth
3 kg
26 cm
38 cm
22 cm
Power
See Section 11.5 Power Supply
Input
Fuse
12VDC (11-13VDC)
UL Listed fuse: 600mA/250Vac (UL 248)
Power consumption
NucleoCounter
Peak
Ready mode
Standby
25 W
2.5 W
2 mW
Operation conditions
Maximum relative humidity 80 percent for temperatures up to 31°C decreasing
linearly to 65 percent relative humidity at maximum 35°C; minimum temperature
15°C.
USB (For image and result
transfer to PC)
USB, version 1.1.
Note: Does not support USB Hubs
RS-232 Printer port
(To be used for printer,
SW-updating and result
transfer to PC)
Baud rate:
Data bits:
Parity:
Stop bits:
Flow control:
19200
8
None
1
None
Other environmental operating conditions for the NucleoCounter SP-100:
Indoor use only
Altitude
Up to 2000 m
Mains supply
Refer to the specifications for the Power supplies in section 11.5 Power supply
Installation Category
II (Refer to UL 61010A-1)
Overvoltage Category
II (Refer to UL 61010A-1)
Pollution Degree
II (Refer to UL 61010A-1)