5 Operation of NucleoCounter SP-100
28
Clean the insertion slit, e.g. with compressed, dry air (refer to section 7 Maintenance of
NucleoCounter SP-100 and Technical Note No. 004 “
How to clean the NucleoCounter”
).
Perform the zero count check again. If the error appears even after several attempts to
clean the insertion slit with compressed air and if necessary the NucleoCounter
Clean
Kit, contact ChemoMetec A/S or the local distributor.
5.12 Setting the Date (F200, only relevant when printer is connected)
The function,
F200, is used to set the date. This feature is only used when a printer is
connected to the printer port. Please, also refer to
Technical Note No. 100
with respect
to setting and using the Date feature.
The date can be set using the
F200 command:
Display Text Type 24
Set date?
Set date?
Set date?
Set date?
2005
2005
2005
2005-
-
-
-02
02
02
02-
-
-
-23
23
23
23
20yy
20yy
20yy
20yy-
-
-
-mm
mm
mm
mm-
-
-
-dd
dd
dd
dd
↵
↵
↵
↵
store
store
store
store
Figure 52. Type in the date and press “↵
↵
↵
↵
” to save the new date or press the “Esc”
button to cancel the operation.
5.13 Setting the Time (F201, only relevant when printer is connected)
The function,
F201, is used to set the Time. This feature is only used when a printer is
connected to the printer port. Please, also refer to
Technical Note No. 100
with respect
to setting and using the Time feature.
The Time can be set using the
F201 command:
Display Text Type 25
Set time?
Set time?
Set time?
Set time?
11:02
11:02
11:02
11:02
hh
hh
hh
hh-
-
-
-mm
mm
mm
mm
↵
↵
↵
↵
store
store
store
store
Figure 53. Type in the time and press “↵
↵
↵
↵
” to save the new time or press the “Esc” to
cancel the operation.
5.14 Resetting the Counter (F30, only relevant when printer is connected)
The function
F30 is used to reset the Counter. The Counter counts the number of
analysis performed by the instrument. The Analysis Number is incremented by 1 each