8 Troubleshooting, Error messages

44

Figure 75. Location and removal of the fuse holder

Disconnect the power supply from the NucleoCounter SP-100. Then remove the fuse
holder by use of a screwdriver (turn CCW) as indicated in

Figure 75

. Inspect the fuse. If

it needs replacement use an UL Listed fuse: 600mA/250Vac (UL 248).

If the NucleoCounter SP-100 does not turn on after replacement of the fuse, or if the
fuse was not broken, please contact ChemoMetec A/S or your local distributor for
further instructions.

8.7

Instrument is not able to finish the analysis

If a PC is connected to the NucleoCounter SP-100 via a USB cable and the PC is not
powered on, then the NucleoCounter SP-100 can have a problem with finishing the
analysis (e.g. the display keeps showing “Analyzing”, see figure below).

Display Text Type 31

Analysing

Analysing

Analysing

Analysing

Figure 76.

The instrument is in the analysis mode

Solution: Remove the USB cable from the instrument or power on the PC.