NIRCal 5.5 Manual, Version A

39

After several calibration optimizations and running the

automatic calibration wizard

, there will be

several calibrations in the project. Sort the calibrations with a click on the Name, Q-Value or e.g. SEP.

NOTE
Keep only the best calibration and delete all others.

2.2.12 Save Calibration - Lifecycle

Save the project and the calibrations to the database. Only approved calibrations according to
Lifecycle will be available in the NIRWare Application Designer. See

Calibration handling

for further

details.

2.2.13 Create a Quantitative Calibration with Cross Validation

The above tutorial is valid for a user selected validation set (VS method).
It is possible to create a quantitative calibration using the

Cross Validation

(CV method), in that case

all spectra should be in the C-Set, the V-Set should be empty.