Save calibration - lifecycle, 12 save calibration - lifecycle – BUCHI NIRCal User Manual
Page 39

Tutorials
NIRCal 5.5 Manual, Version A
39
After several calibration optimizations and running the
automatic calibration wizard
, there will be
several calibrations in the project. Sort the calibrations with a click on the Name, Q-Value or e.g. SEP.
NOTE
Keep only the best calibration and delete all others.
2.2.12 Save Calibration - Lifecycle
Save the project and the calibrations to the database. Only approved calibrations according to
Lifecycle will be available in the NIRWare Application Designer. See
Calibration handling
for further
details.
2.2.13 Create a Quantitative Calibration with Cross Validation
The above tutorial is valid for a user selected validation set (VS method).
It is possible to create a quantitative calibration using the
Cross Validation
(CV method), in that case
all spectra should be in the C-Set, the V-Set should be empty.