BUCHI NIRCal User Manual
Page 143

Chemometrics
NIRCal 5.5 Manual, Version A
143
Lifecycle: Copy
A copy of the active calibration can be created with this LC function. There is a request for calculation
in case after a modification the calculation is not yet made.
The copy is automatically saved in the DB and set to active with the status: Created.
The copy has the same GUID, as the original calibration, but a higher version number.
For the copy of a calculated calibration the Prediction Protocols are available.
Each calibration can be copied.
The copied calibration can be edited, modified and used later -after the necessary steps of saving- for
the application.
Lifecycle: Next
To put a created calibration to the approved status, this LC function is used. If the calibration is not
yet calculated, there is a request for calculation.
NOTE
It is not possible to put the first master calibration to the next status.
After the necessary calculation the calibration is stored and it`s quality is tested. The Q-Value default
limit is:
0.8 for qualitative and
0.6 for quantitative calibrations.
It is possible to put a calibration to a higher status with lower Q-value.
NOTE
Do not use a calibration with outliers!
Lifecycle: Delete
Calibrations, which are created but not yet stored in the DB, can be deleted by selecting them in the
NIR-Explorer and pressing "Delete".
Calibrations, which are already stored in the DB but not used for application, can be deleted only
with: Lifecycle / Delete.
The question about deleting should be answered before: