Cv secv, Cv spectra residuals, Cv v-set press – BUCHI NIRCal User Manual
Page 175: C-set bcap-pres, 20 cv secv, 21 cv spectra residuals, 22 cv v-set press, 23 c-set bcap-pres

Chemometrics
NIRCal 5.5 Manual, Version A
175
3.18.20
CV SECV
Description
CV SECV: Standard Error Cross Validated
Use
The CV selects the number of secondary PCs, where CV SECV has a
minimum.
Method
PCR / PLS
Matrices ID
72
Tip
The secondary PCs selection is indicated with red color.
Details
The first local minimum of CV SECV is taken.
Formula
Related Topic
SEE
3.18.21
CV Spectra Residuals
Description
The spectra residual of the spectra in the V-Set during CV.
Use
Compare the CV spectra residual with the residual of final calibration to find
out possible outliers.
Method
PCR / PLS
Matrices ID
71
Tip
Big CV spectra residual indicates outlier.
Details
The number of primary PCs are used for the calculation.
Related Topic
Residuals
,
Primary PCs
3.18.22
CV V-Set PRESS
Description
CV V-Set PRESS is the Predicted Residual Error Sum Square for the
samples in case they are in the V-Set of the calibration.
Use
To estimate the number of PCs. Compare it with C-Set PRESS of the final
calibration.
Method
PCR / PLS
Matrices ID
69
Tip
Optimal number of PC, where CV V-Set PRESS has the minimum value.
Details
Shows the CV C-Set PRESS in function of the number of PCs.
Related Topic
C-Set PRESS
,
CV C-Set PRESS
3.18.23
C-Set BCAP-PRES
Description
PRES function of the calibration Set comparable with the DOS based BCAP
Software.
Use
To estimate the number of factors
Method
PCR / PLS
Matrices ID
21
Tip
Details
Shows the C-Set PRES (Predicted Error Sum) in function of the number of
PCs.
Related Topic
V-Set BCAP Pres