7 Maintenance and repairs

73

B-395 Pro

Operation Manual, Version C

Figure 7-1: Cleaning procedure of the nozzle

Take a syringe containing air on top and water on the bottom.

Push the air through the nozzle (left figure).

Push the water through the nozzle immediately afterwards (right figure).

Examine the nozzle tip under a stereoscopic microscope to make sure the passage is
clear and clean.

NOTE
If lipophilic immobilization solutions were used, then use appropriate solvents for cleaning. Do not use
acid solution for alginate, as this would create a precipitate.

7 .4 .2

Cleaning a clogged nozzle

Unscrew the nozzle. Pass air or water through the nozzle as shown in figure 7-1.

If the nozzle tip is not clear, soak the nozzle in water, the appropriate solvent, in 1N NaOH or 1N
sulfuric acid (never use HCl) according to the encapsulation mixture for 1 hour at room temperature,
with periodic agitation. Sonication of full stainless steel nozzles is also a helpful procedure. Wear
appropriate protection equipment. Rinse with distilled water, with air and let dry.

Examine the nozzle tip under a stereoscopic microscope to make sure the passage is clear and clean.

NOTE
If lipophilic immobilization solutions were used, then use appropriate solvents for cleaning. Do not use
acid solution for alginate, as this would create a precipitate.

7 .4 .3

Cleaning the reaction vessel and the other vessels

Disassemble the reaction vessel. However the magnet holder should not be disassembled!
Disassemble the liquid transfer system.
Wash all parts, except the air filters, with a mild detergent solution, 0.01N NaOH or 0.01N sulfuric acid
(never use HCl) as appropriate.
Rinse thoroughly with hot water, then with distilled water and let dry.