1 menu for syringe pump calibration – BUCHI Encapsulator B-395 Pro User Manual

Page 42

background image

6 Operation

42

B-395 Pro

Operation Manual, Version C

Pressing on the (+) and (–) buttons will change the pump-
ing rate. Pressing “Esc” will return you to the start menu
and the set value will be kept. Pressing the “cal mL/min”
button, while the pump is running, will open screen 6-11
and allows you to calibrate the current syringe. If the pump
is stopped, the screen will ask you to select a calibrated
syringe.

Screen 6-10: Speed regulation of the syringe pump

6 .5 .1

Menu for syringe pump calibration

Select appropriate syringe volume by pressing the cor-
responding button. You are forwarded to screen 6-12
(if the pump is running) or to the main screen (if the pump
is stopped).

Screen 6-11: Syringe pump calibration - selecting syringe
type

Pressing button “on” starts timer. The timer counts down
for 1 minute from 60 to 0 sec. During this time the liquid
from the jet is collected in a pre-weighted vessel. The
three last seconds are announced by a short tone. One
second later the syringe pump stops and you are forward-
ed to screen 6-13.

Screen 6-12: Syringe pump calibration - timer