2 selecting a calibrated syringe – BUCHI Encapsulator B-395 Pro User Manual
Page 43

6 Operation
43
B-395 Pro
Operation Manual, Version C
Pressing on the (+) and (–) buttons for entering the liquid
pumped for 1 min. Then press button “store”.
You are forwarded to screen 6-14.
Screen 6-13: Syringe pump calibration - setting the pumped
liquid
Press “Yes” to store the values. You are forwarded to the
start menu. The syringe pump will now run with the new
calibration after pressing the “on” button of the pump
control.
Screen 6-14: Syringe pump calibration - store value
6 .5 .2
Selecting a calibrated syringe
Stop the pump. Press on screen 6-10 the “cal mL/min” button. You are forwarded to screen 6-15.
Select the appropriate syringe by pressing on the item.
The syringe is stored and you are forwarded to the start
menu.
NOTE
This screen is only accessible from the start menu, if the
pump is stopped.
Sreen 6-15: Selecting the syringe size