Emissions and Immunity Information

Radiated and Conducted Emissions were tested on the Elgar power source. It complies with the
EN55022 Class B limits, and FCC part 15 which as same limits as EN55011 Group 1, Class B.
Harmonics and Flicker testing was not tested, as the EUT is operated at 120V, 60Hz, which is
beyond the scope of those standards.

Guidance and Manufacturer’s Declaration - Emissions

The Image Capture System is intended for use in the electromagnetic environment specified below. The

customer or user of the Image Capture System should ensure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF Emissions

CISPR11

Class B

Group 1

The Image Capture System uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.

RF Emissions

CISPR22, FCC

Class B

The Image Capture System uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.

Harmonics

IEC 61000-3-2

N/A

For use in domestically and in Canada.

Beyond the scope of the standard.

Flicker

IEC 61000-3-3

N/A

For use in domestically and in Canada.

Beyond the scope of the standard.

The Image Capture System is suitable for use in all
establishments, including domestic, and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.