Factory default settings – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 79

Directions for Use
Chapter 7 Reference
75
Factory default settings
Function
Value
Blood pressure measurement units
mmHg
Patient type
Adult
Automatic NIBP Interval
15 minutes
MAP (Adult and Pediatric)
Disabled
MAP (Neonate)
Enabled
NIBP Adult High Systolic Alarm
220 mmHg
NIBP Adult Low Systolic Alarm
75 mmHg
NIBP Adult High Diastolic Alarm
110 mmHg
NIBP Adult Low Diastolic Alarm
35 mmHg
NIBP Adult High MAP Alarm
120 mmHg
NIBP Adult Low MAP Alarm
50 mmHg
NIBP Pediatric High Systolic Alarm
145 mmHg
NIBP Pediatric Low Systolic Alarm
75 mmHg
NIBP Pediatric High Diastolic Alarm
100 mmHg
NIBP Pediatric Low Diastolic Alarm
35 mmHg
NIBP Pediatric High MAP Alarm
110 mmHg
NIBP Pediatric Low MAP Alarm
50 mmHg
NIBP Neonatal High Systolic Alarm
120 mmHg
NIBP Neonatal Low Systolic Alarm
50 mmHg
NIBP Neonatal High Diastolic Alarm
70 mmHg
NIBP Neonatal Low Diastolic Alarm
30 mmHg
NIBP Neonatal High MAP Alarm
80 mmHg
NIBP Neonatal Low MAP Alarm
35 mmHg
SpO
2
Adult High Alarm
---
SpO
2
Adult Low Alarm
85%
SpO
2
Pediatric High Alarm
---
SpO
2
Pediatric Low Alarm
85%
SpO
2
Neonatal High Alarm
95%
SpO
2
Neonatal Low Alarm
85%
Temperature Scale
F (Fahrenheit predictive)
Pulse Rate Adult High Alarm
120 beats per minute
Pulse Rate Adult Low Alarm
50 beats per minute
Pulse Rate Pediatric High Alarm
150 beats per minute
Pulse Rate Pediatric Low Alarm
50 beats per minute
Pulse Rate Neonatal High Alarm
200 beats per minute
Pulse Rate Neonatal Low Alarm
100 beats per minute
Pulse Tone Volume
03
Print Control
Batch
Time-of-day Display
24-hour