Monitor specifications, Performance – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 70

66
Chapter 7 Reference
Welch Allyn Vital Signs Monitor 300 Series
Monitor specifications
The monitor is an FDA Class II non-critical device and, according to Council Directive 93/
42/EEG, Annex IX, a Class IIB device.
Performance
Characteristic
Specification
Notes
General
Recovery from Defibrillation
Discharge
Per IEC 60606-2-30:1999(E) The monitor returns to normal function within 1 minute after
the discharge of a cardiac defibrillator. (Actual recovery time = 0 seconds)
WARNING Always disconnect the SpO
2
sensor cable from the monitor
before defibrillating the patient.
NIBP
Cuff Pressure Range
0 to 300 mmHg (0 to 40 kPa)
Initial Cuff Inflation
Factory Default
Adult
160 mmHg (21.3 kPa)
Pediatric
120 mmHg (16 kPa)
Neonate
90 mmHg (12 kPa)
Blood Pressure Accuracy
•
Blood pressure measurement meets or exceeds ANSI/AAMI
SP10: 2002 for manual, electronic, or automated
sphygmomanometers.
•
Blood pressure accuracy is validated for pressure measurement
using the upper arm only, with the patient seated.
•
Blood pressure is validated against manual auscultatory readings
for adults and children older than 3 years.
•
Blood pressure is validated against intra-arterial readings for
children 3 years and younger.
Blood Pressure
Determination Time
20 - 45 seconds typical; 165 seconds maximum.
Overpressure Cutoff
295 to 330 mmHg (39.3 to 44 kPa)
Systolic Range
Adult
30 to 260 mmHg (4 to 34.5 kPa)
Pediatric
30 to 160 mmHg (4 to x 21.3 kPa)
Neonate
25 to 120 mmHg (3.3 to 16 kPa)
Diastolic Range
Adult
20 to 235 mmHg (2.7 to 31.3 kPa)
Pediatric
15 to 130 mmHg (2 to 17.3 kPa)
Neonate
10 to 105 mmHg (1.3 to 14 kPa)
MAP
Adult
20 to 255 mmHg (2.7 to 30 kPa)
Pediatric
15 to 140 mmHg (2 to 18.7 kPa)
Neonate
10 to 110 mmHg (1.3 to 14.7 kPa)
Pulse Rate Range
30 to 220 bpm
Pulse Rate Accuracy
± 3 bpm or 3%
Note
NIBP specifications are tested by Welch Allyn using the BIO-TEK BP Pump NIBP
Monitor Tester.