beautypg.com

Other environmental specifications – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual

Page 77

background image

Directions for Use

Chapter 7 Reference

73

Other environmental specifications

Recommended separation distances between portable and mobile RF communications equipment and the

Vital Signs Monitor 300 Series monitor

The Vital Signs Monitor 300 Series monitor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Vital Signs Monitor 300 Series monitor can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Vital Signs Monitor 300 Series monitor as recommended below, according to the
maximum output power of the communications equipment.

Rated maximum output

power of transmitter

W

Separation distance according to frequency of transmitter

(meters)

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 1.2

800 MHz to 2.5 GHz

d = 2.3

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (w) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

P

P

P

Characteristic

Specification

Operating Temperature

SpO2, NIBP, pulse rate, and
temperature measurement

+10

C to +40 C

+50

F to +104 F

Operating altitude

-170 m to +4877 m
-557 ft to + 16,000 ft

Shipping altitude

-615 m to 12,300 m
-2000 ft to 40,000 ft

Storage temperature

-20

C to +50 C

-4

F to +122 F

Relative humidity

15% to 90% (non-condensing)