Error legend, Footer – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 63

Directions for Use
Chapter 5 Reviewing patient data
59
Pulse rate
The pulse rate notation does not include error information.
SpO
2
SpO
2
notation can include error information:
NIBP
NIBP notation can include error information:
Error legend
If any error indicators appear in the printed data, the monitor prints an error legend. The
error legend contains a two-line header followed by a brief explanation of each error type
encountered in the data. The illustration below shows the error legend for a batch of data
containing at least one temperature (P) error, at least one NIBP C04 error, and no
other errors.
Footer
The printout footer consists of a line containing the monitor serial number, a line
containing the monitor software version, two separator lines, and four blank lines:
C7
Malfunctioning sensor.
--
Sensor error.
C01
NIBP cycle cancelled by operator.
C02
Unable to calibrate.
C03
Cuff inflation to rapid.
C04
Excessive cuff inflation time.
C05
Excessive noise or air leak.
C06
Measurement out of range.
C10
Cuff overpressure.
| Error Codes: |
| ------------ |
| (P) Loss of tissue contact |
| C04 Excessive inflation time |
| |
| Unit S/N: JA736455 |
| S/W Ver.: 1.00.00 00005 |
| -------------------------------- |
| -------------------------------- |
| |
| |
| |
| |