Directions for Use

Chapter 4 Alarms and alerts

51

Nurse call

The monitor can be connected to a Nurse Call system through a customized cable that
connects to the Nurse Call connector. When the cable is connected and operational, the
monitor immediately notifies the Nurse Call system when a patient alarm occurs.

Alarm limits for vital signs

Vital sign

Low limit (default)

High limit (default)

Resolution

Systolic (mmHg)

Neonatal

30 - 110 (50)

35 - 115 (100)

5 mmHg (0.5 kPa)

Pediatric

35 - 150 (75)

40 - 155 (145)

5 mmHg (0.5 kPa)

Adult

35 - 250 (75)

40 - 255 (220)

5 mmHg (0.5 kPa)

Diastolic (mmHg)

Neonatal

15 - 95 (30)

20 - 100 (70)

5 mmHg (0.5 kPa)

Pediatric

20 - 120 (35)

25 - 125 (100)

5 mmHg (0.5 kPa)

Adult

25 - 225 (35)

30 - 230 (110)

5 mmHg (0.5 kPa)

Mean arterial pressure (mmHg)

Neonatal

15 - 100 (35)

20 - 105 (80)

5 mmHg (0.5 kPa)

Pediatric

20 - 130 (50)

25 - 135 (110)

5 mmHg (0.5 kPa)

Adult

25 - 245 (50)

30 - 250 (120)

5 mmHg (0.5 kPa)

Pulse rate (beats per minute)

Neonatal

35 - 210 (100)

40 - 215 (200)

5 bpm

Pediatric

35- 210 (50)

40 - 215 (150)

5 bpm

Adult

35 - 210 (50)

40 - 215 (120)

5 bpm

SpO

2

(%)

Neonatal

50 - 98 (85)

51 - 99 (95)

1%

Pediatric

50 - 98 (85)

51 - 99 (---)

1%

Adult

50 - 98 (85)

51 - 99 (---)

1%

Nurse Call Cable

Connector