48

Chapter 4 Alarms and alerts

Welch Allyn Vital Signs Monitor 300 Series

Recoverable SpO

2

alert—escalated

An SpO

2

equipment alert is always escalated immediately to the level of a patient alarm if

it occurs when both of the following conditions exist:

SpO

2

monitoring has begun and the monitor has recorded an SpO2

measurement

An SpO

2

or Pulse Rate alarm limit has been set

See

“Responding to a patient alarm”

on page 47.

Recoverable NIBP alert—escalated

An NIBP equipment alert is escalated to the level of a patient alarm whenever both of the
following conditions exist:

Two consecutive NIBP equipment alerts occur while the monitor is taking
automatic NIBP/PR measurements

Alarms are enabled

See

“Responding to a patient alarm”

on page 47.

Nonrecoverable alerts

When the monitor detects a nonrecoverable equipment problem, it does the following:

Displays an error code in the SYS window, and shuts off the display to all
other windows

Stops patient monitoring

Stops the pump and opens the air valve

Activates the Nurse Call relay (if connected)

Produces an audible tone

Shuts down as soon as

is pressed or one minute has elapsed

Response

For both recoverable and nonrecoverable equipment alerts, respond as follows:

1.

Press

to immediately silence the alert tone.

2.

Determine what caused the alert and correct the problem.

Note

For information about a battery alert, see

“Battery low warning”

on page 65 and

“Battery failure”

on page 65.