Chapter 5 Reviewing patient data

Welch Allyn Vital Signs Monitor 300 Series

Date

Time

The recording time is printed for each record, in the form hh:mm (using a 24-hour clock),
at the beginning of the first line of the record.

In the illustration below, the first record (21:45) contains systolic and diastolic pressure,
MAP, pulse rate, and SpO2 readings; the second record (21:52) contains systolic and
diastolic pressure, MAP, pulse rate, and temperature; and the third record (22:12) contains
only a temperature reading. When temperature is the only recorded vital sign, the time is
printed on the same row as the temperature reading.

Range and alarm flags

Flag characters (

 and ) indicate a data value that falls outside of a defined range or

violates a programmed alarm threshold. In the illustration above, 69

 indicates a diastolic

pressure value below the programmed alarm limit, and 110.0F

indicates a temperature

value above the monitor’s temperature measurement limit.

Temperature

The temperature notation includes information about the type and location of the
measurement, and can include error notation:

| DD-MMM-YYYY |

| Time Sys Dia MAP PR SpO2 |
| ---- mmHg ---- BPM % |
| --------------------------------- |
| 21:45 125 69

90 72 98 |

| 21:52 125 69

90 72 98 |

| Temp 99.2F NOR/ORL |
| 22:12 Temp 110.0F

NOR/ORL |

Type

F

Fahrenheit

C

Celsius

NOR

Normal (predictive)

MON

Monitored

Location

OrL

Oral

(Oral probe)

AP

Axillary Pediatric

(Oral probe)

AA

Axillary Adult

(Oral probe)

rEC

Rectal

(Rectal probe)

Error

(P)

Discontinuity in tissue contact during the measurement
period. The displayed temperature measurement is not
necessarily accurate.

C20

Broken or missing temperature probe.