Chapter 3 Patient monitoring

Welch Allyn Vital Signs Monitor 300 Series

Setting the temperature measurement type

The monitor, if configured with the temperature option, can provide both predictive and
monitored temperature measurements.

A predictive measurement is a one-time measurement that takes only a few seconds. It
results in a single temperature reading which is displayed at the end of the brief
measurement period. The monitor sounds three short tones to indicate the end of a
predictive measurement.

A monitored measurement is a continuous temperature monitoring, used when the
situation prevents accurate predictive measurement. For oral and rectal measurements,
three minutes of monitoring is recommended. For axillary measurements, five minutes of
monitoring is recommended.

During a monitored measurement, the temperature is displayed dynamically throughout
the measurement period. Unlike a predictive measurement, the monitor does not indicate
the end of any elapsed time for a monitored measurement.

To select the temperature measurement type:

1.

Press

repeatedly until TEMP MODE appears in the display window.

2.

Press

or

to cycle to the option you wish to select:

3.

To set the temperature measurement type and return the monitor to normal
operation, do nothing for 10 seconds or press any button other than

or

. If you

press a function button (such as

), the monitor returns to normal operation with that

function (

) activated.

Loading a probe cover

1.

Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.

2.

Insert the probe into a probe cover and press the probe handle down firmly. The
probe handle moves slightly to engage the probe cover.

WARNING Do not exceed the recommended measurement periods of three
minutes for oral and rectal measurements and five minutes for axillary
measurements.

Fahrenheit predictive

Celsius predictive

Fahrenheit monitored

Celsius monitored

ºF

ºF

ºC

ºC

ºF

ºF

M

M

M

M

ºC

ºC