Directions for Use

Chapter 7 Reference

71

Guidance and manufacturer’s declaration—electromagnetic immunity

The Vital Signs Monitor 300 Series monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment—

guidance

Electrostatic discharge
(ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

±2 kV for power supply
lines
±1 kV for input/output
lines

±2 kV for power supply
lines
±1 kV for input/output
lines

Mains power quality should be that of a
typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV differential mode
±2 kV common mode

±1 kV differential mode
±2 kV common mode

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11

<5% U

t

(>95% dip in U

t

)

for 0.5 cycle
40% U

t

(60% dip in U

t

)

for 5 cycles
70% U

t

(30% dip in U

t

)

for 25 cycles
<5% U

t

(>95% dip in U

t

)

for 5 sec

<5% U

t

(>95% dip in U

t

)

for 0.5 cycle
40% U

t

(60% dip in U

t

)

for 5 cycles
70% U

t

(30% dip in U

t

)

for 25 cycles
<5% U

t

(>95% dip in U

t

)

for 5 sec

Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the VSM 300
Series monitor requires continued
operation during power mains
interruption, it is recommended that the
monitor be powered from an
uninterruptible power supply or a battery.

Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.

Note

U

t

is the AC mains voltage prior to application of the test level.