beautypg.com

Map measurement enable and disable – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual

Page 27

background image

Directions for Use

Chapter 2 Setup

23

MAP measurement enable and disable

Default setting: MAP ENABLED for neonate; MAP DISABLED for adult and pediatric.

1.

Depress

for 3 seconds. SET HOUR XX appears in the message window.

2.

Press

repeatedly until MAP ENABLED or MAP DISABLED appears in the

display window.

3.

Press

or

to enable or disable MAP measurement.

4.

To select the displayed state and return to normal operation, either do nothing for 10
seconds or press any button other than

or

. If you press a function button (such

as

), the monitor returns to normal operation with that function (

) activated.

For information about MAP measurements, see

“Patient monitoring”

on page 29.

Note

If you change the MAP enabled/disabled setting, refer to

“How changing the

patient type affects MAP defaults”

on page 34.

SYS

DIA

SpO

2

%

/min

ºF

ºF

mmHg

MAP DISABLED

SYS

DIA

SpO

2

%

/min

ºF

ºF

mmHg

MAP ENABLED