Directions for Use
Chapter 7 Reference
67
Physical
Characteristic
Specification
Notes
Temperature
Temperature Measurement
Range
80
F to 110 F
26.7
C to 43.3 C
Temperature Calibration
Accuracy
± 0.2
F (± 0.1 C); meets or exceeds ASTME1112-00; EN12470-3:2000
Temperature Determination
Time
predictive
(typical)
oral
axillary
rectal
4 seconds
10 seconds
15 seconds
SpO
2
Characteristic
Specification
Notes
All sensors have a measurement
range of 70% - 100%. SpO2 complies
with EN ISO 9919:2005.
Pulse Rate Range
25 to 240 bpm
Pulse Rate Accuracy
± 3 bpm
± 5 bpm
No motion
Motion
± 3 digits
Low perfusion
SpO
2
Accuracy
70-100% ± 2 digits
Adult and pediatric
70-100% ± 3 digits
Neonate
Pulse Oximetry (SpO
2
)
Saturation Range
1% to 100%
Electrosurgery interference
suppression
Yes
Sensor compatibility
For a complete list of compatible sensors refer to
Masimo
For probe/sensor compliance to EN ISO 9919:2005, see the Masimo directions for use.
Sensor lights (Masimo)
Red wavelength
Red wavelength (toe clip)
Infrared wavelength
Infrared wavelength (toe
clip)
15 mW at 50 mA pulsed
660 nm
663 nm
905 nm
880 nm
Risk mitigation: Software in this device is developed under the FDA guidance of Part 802.3 design controls and the
international standard EN 14971 risk management. These mitigate and minimize any risk associated with potential
software errors to a level as low as reasonably possible (ALARP).
Note
SpO
2
specifications are published by the SpO
2
component manufacturer and
tested by Welch Allyn using the BIO-TEK Index 2 SpO
2
simulator.
Characteristic
Specification
Dimensions
height
6.6 inches
16.8 cm
width
10.0 inches
25.4 cm
depth
6.0 inches
15.2 cm
Weight
Approximately 5.4 lbs (2.4 kg)
Color, Temperature Probe
Oral/Axillary
Blue
Rectal
Red