4 keypad functions – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual
Page 9

Counterpart User Manual - Introduction
3
1.4
Keypad Functions
Key
Function
POWER
Turns the
Counterpart
unit on/off.
NOTE: If the PC1 jumper is set to SW, the
POWER
button must be used to turn the unit on and off. If the PC jumper is set
to ON, the unit will automatically power on when it’s plugged in and the only way to turn it off is to unplug power.
Enters Menu mode, allowing configuration if the Audit jumper is in the “ON” position. Also used as an “escape” key in Menu
mode.
Sets the current gross weight to zero, provided the amount of weight to be removed or added is within the specified zero
range and the scale is not in motion. The zero band is defaulted to 1.9% of full scale, but can be configured for up to 100%
of full scale.
NOTE: 1.9% is required for legal for trade applications.
Used to enter a known unit weight of a part.
Sends “on-demand” serial information out the serial port, provided the conditions for standstill are met.
Performs one of several predetermined Tare functions dependent on the mode of operation selected in the TAREFN
parameter. To view a stored tare, see Section 1.5.
Also acts as the
ENTER
key
Allows the display to change between gross mode and net mode.
This is used for computing the unit weight by sampling.
Press the
SAMPLE
key after placing a sample on the platform, or after using the numeric keypad to enter the sample
size.
The default sample size is 10.
This switches the weight display to an alternate unit. the alternate unit is defined in the Setup Menu and could be kg, g, lb,
oz.
Clear key. Used to backspace on entries
Alpha-Numeric
keypad (see
Figure 1-1)
Used to enter values directly. Press keys for small and upper case alpha characters.
Table 1-2. Keypad Functions
MENU
SETUP
ZERO
UNIT
WEIGHT
TARE
B/N
GROSS
NET
SAMPLE
Pieces
UNITS
TOGGLE