9 indicator operations – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual
Page 11

Counterpart User Manual - Introduction
5
1.9
Indicator Operations
Basic
Counterpart
operations are summarized below.
Toggle Gross/Net
Press the
GROSS/NET
key to switch the display mode
from gross to net or from net to gross. If a tare value
has been entered or acquired, the net value is the gross
weight minus the tare. If no tare has been entered or
acquired, the display remains in gross mode.
Gross mode and Net mode is indicated by the
annunicator on the display.
Toggle Units
Press the
UNITS TOGGLE
key to switch between
primary, secondary, and tertiary units.
Zero Scale
1. While in gross mode, remove all weight from
t h e s c a l e a n d w a i t f o r t h e s t a n d s t i l l
annunicator (
).
2. Press the
ZERO
key. The zero annunicator is
displayed to indicate the scale is zeroed.
Acquire Tare
1. Place a container on the scale and wait for the
standstill annunicator (
).
2. Press the
TARE
key to acquire the tare weight
of the container.
3. The
Net
annunicator will be enabled.
Remove Stored Tare Value
1. Remove all weight from the scale and wait for
the standstill annunicator (
).
2. Press the
TARE
key. The display shifts to gross
weight and the Gross annuciator is displayed.
Keyed Tare
1. Key in the value of the container.
2. Press the
TARE
key.
3. The
Net
annunciator is displayed.
Press the
key to print either the Gross, Net,
Total or Count format.
The default format is a 2.25 x 4.5 inch label.
1. Wait for standstill annunciator (
).
2. Press the
key to send data to the serial
port.
If the standstill annunciator is not present, the
key command is executed if the scale comes out of
motion within three seconds.
Set Date and Time
1. If the softkey is enabled, push the softkey.
The current value is displayed.
2. To change, use the Up/Down arrows to
highlight the value to change.
3. Press the
TARE
key to save and exit.
Set Serial Scale (Scale 3)
1. Select Port 1 - Indust Scale
2. Match the baud rates. Upon doing this it
automatically becomes Scale 3.