Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual
Page 62

56
Counterpart User Manual
REG.BASE
Zero base for overload calculation
CALIB, SCALE
REG.PRTMOT
Allow print while in motion
NO, YES
REG.PRINTPT
Add PT to keyed tare print
NO, YES
REG.SNPSHOT
Display or Scale weight source
DISPLAY, SCALE
SD
Set date
MMDDYY, DDMMYY, YYMMDD, or YYDDMM. Enter six-digit
date using the year-month-day order specified for the
DATEFMT parameter, using only the last two digits of the year.
ST
Set time
hhmm (enter using 24-hour format)
TIMEFMT
Time format
12HOUR, 24HOUR
TIMESEP
Time separator
COLON, COMMA
Command
Description
Values
GFMT.FMT
Gross demand print format string
See Section 7.0 on page 58 for information about demand print
format strings.
NFMT.FMT
Net demand print format string
ACC.FMT
Accumulator print format string
CFMT.FMT
Count format
TOTAL.FMT
Total Format
HDRFMT1
Header 1 Format
PALFMT.FMT
Pallet Format
HDRFMT2
Header 2 Format
Table 6-9. PFORMT Serial Commands
Command
Description
Values
DIO#b
Digital input function
OFF, ZERO, NT/GRS, TARE, UNITS, PRINT,
ACCUM, TIMDATE, ESC, CLEAR, DSPTAR,
IDKEY, KEY0–KEY9, KEYDP, ENTER, NAVUP,
NAVDN, NAVLFT, NAVRGT, KBDLOC, HOLD,
CLRCN, GROSS, NET, PRIM, SEC, CLRTAR,
CLRACC, TRIGGER
DIO#b
Digital output function
OVER, UNDER, ACCEPT, ZEROBD
Digital inputs and outputs are specified by bit number
Table 6-10. DIG I/O Serial Commands
Command
Description
Value
GFMT.PORT
Set the port for each of these commands
2
NFMT.PORT
ACC.PORT
CFMT.PORT
TOTAL.PORT
PALFMT.PORT
Table 6-11. Select Port Serial Commands
Command
Description
Values
Table 6-8. FEATURE Serial Commands (Continued)