Ables 6-6 – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 1.0 User Manual
Page 60

54
Counterpart User Manual
Command
Description
Values
SC.GRADS#n
Graduations
1–100000
SC.ZTRKBND#n
Zero track band
0, 0–100
SC.ZRANGE#n
Zero range
1.900000, 0–100
SC.MOTBAND#n
Motion band
1, 0–100
SC.SSTIME#n
Standstill time
1–65535
SC.OVRLOAD#n
Overload
FS+2%, FS+1D, FS+9D, FS
SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n
Digital filtering
1, 2, 4, 8, 16, 32, 64, 128, 256
SC.DFSENS#n
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
SC.DFTHRH#n
Digital filter cutout threshold
NONE, 2D, 5D, 10D, 20D, 50D, 100D, 200D, 250D
SC.THRESH#n
Zero threshold
0–9999999
SC.SMPRAT#n
Sample rate
30HZ, 60HZ, 120HZ, 240HZ, 480HZ, 960HZ
SC.PWRUPMD#n
Power up mode
GO, DELAY
SC.TAREFN#n
Tare function
BOTH, NOTARE, PBTARE, KEYED
SC.PRI.DECPNT#n
Primary units decimal position
8.888888, 88.88888, 888.8888, 8888.888, 88888.88,
888888.8, 8888888, 8888880, 8888800
SC.PRI.DSPDIV#n
Primary units display divisions
1D, 2D, 5D
SC.PRI.UNITS#n
Primary units
lb, kg, g, oz, OFF
SC.ACCUM#n
Accumulator enable
ON, OFF
SC.WZERO#n
Zero calibration
—
SC.WVAL#n
Test weight value
test_weight_value
SC.WSPAN#n
Span calibration
—
SC.WLIN.F1–
SC.WLIN.F5#n
Actual raw count value for linearization
points 1–5
0–16777215
SC.WLIN.V1
SC.WLIN.V5#n
Test weight value for linearization points
1–5
0.000001–9999999
SC.WLIN.C1–
SC.WLIN.C5#n
Calibrate linearization points 1–5
—
SC.LC.CD#n
Deadload coefficient
—
SC.LC.CW#n
Span coefficient
—
SC.LC.CZ#n
Temporary zero
—
SC.REZERO#n
Rezero
—
SC.SEC#n
Secondary units
lb, kg, g, oz, OFF
SC.TER#n
Tertiary units
lb, kg, g, oz, OFF
Table 6-6. SCALES Serial Commands
Command
Description
Values
EDP.BAUD#p
Port baud rate
1200, 2400, 4800, 9600, 19200, 28800, 38400, 57600, 115200
EDP.BITS#p
Port data bits/parity
8NONE, 7EVEN, 7ODD
EDP.STOPBITS#p
Port stop bits
1, 2
EDP.PRNMSG#p
Print message
ON, OFF
Table 6-7. SERIAL Port Serial Commands