beautypg.com

Archived 4/2/10, Icnirp (general public), Safety code 6 (canada) exposed workers – ETS-Lindgren HI-4460 Graphical Readout (Archived) User Manual

Page 64

background image

64

| Standards

ICNIRP (General Public)

Frequency (f) in same units as defined in first column.

Power

Density

(mW/cm

2

)

E-Field

(V/m)

H-Field

(A/m)

E-Field

H-Field

0-1Hz

C

C

3.2 x 10

4

1Hz-8Hz

C

10,000

3.2 x 10

4

/f

8Hz-25Hz

C

10,000

4,000/f

0.025kHz-0.8kHz

C

250/f

4/f

0.8kHz-3kHz

C

250/f

5

3kHz-150kHz

C

87

5

0.15MHz-1MHz

C

87

0.73/f

1MHz-10MHz

C

87/f

1/2

0.73/f

10MHz-400MHz 0.2

28

0.073

400MHz-2000MHz f/2000

1.375f

1/2

0.0037f

1/2

2GHz-300GHz 1

61

0.16

Safety Code 6 (Canada) Exposed Workers

Frequency (f) in MHz.

Power

Density

(mW/cm

2

)

E-Field

(V/m)

H-Field

(A/m)

(MHz) E-Field

H-Field

0.003-1

C

600

4.9

1-10

C

600/f

4.9/f

10-30

C

60

4.9/f

30-300 1

60

0.163

300-1500 f/300

3.54f

1/2

0.0094f

1/2

1500-40,000 5 137

0.364

Archived 4/2/10