8.0 Logging

The logging feature is used to gather readings from the probe and save them for
later review. Each set of logged data contains a log file header. This header
includes the setup information for a particular set of data. A new log file is
created any time the setup is changed. Because a change in the setup is
required for each set of scheduled data, a new log file is created each time it is
used. For single point and Periodic logging, the data is appended to the end of
the log file. When using continuous logging, a new log file is created every time
logging is stopped and started again.

Four menus are available exclusively to the logging screens: SETUP, REVIEW,
UPLD, and CLEAR.

SETUP Menu

The SETUP menu is used to select
parameters for logging operations.
The four possible logging operations
are single point, continuous, periodic,
and scheduled. The submenu log is
available to choose the items to be
included in the log file.