beautypg.com

Archived 4/2/10 – ETS-Lindgren HI-4460 Graphical Readout (Archived) User Manual

Page 44

background image

44

| Logging

START DATE—Defines the date to start logging. Enter any valid date

format.

DURATION—Defines the length of time to log. Enter any time between

00:00:00 and 23:59:59.

For scheduled logging to begin, the LOG key must be pressed before
or during the time of the intended recording. You may turn the HI-4460
off after the LOG key is pressed; it will automatically turn on at the
defined start time.

Log Moving Average

The log moving average is computed using the log interval and the moving
average size values. A log moving average buffer is dynamically set up to hold
the number of logged values that can be taken in the moving average size time.

For example, for a log interval of 1/second and a moving average size of 6
minutes, the moving average is the average of the last 360 logged values, if they
exist. If there have not been 360 logged values, then zeros will be inserted as
place holders and used to calculate the average. When the 361st value is taken,
it will replace the first, and so on.

REVIEW Menu

Views the data in a log file.

Archived 4/2/10