Logging

|

41

C

ONTINUOUS

L

OGGING

Logs continuously at the rate
selected. Press the LOG key to start
logging and press it again to stop.

LOG INTERVAL—Sets the interval between logged values. Valid settings

include 1/minute to 2/second.

MOV AVG SIZE—Sets the time interval used to calculate the log moving

average. Select a value between 0.0 and 99.9 using the number picker
function. If 0.0 is displayed or chosen, there is no log moving average, so it
will not be displayed or included in the log file.

TRIG LO—Selects the lower data logging value. Logging begins when a

field level below this value exists.

TRIG HI—Selects the upper data logging value. Logging begins when a field

level above this value exists.

The setup values for LOG INTERVAL, MOV AVG SIZE, TRIG LO, and
TRIG HI do not change when the logging function is changed. If new
values are desired, they must be changed.