Archived 4/2/10 – ETS-Lindgren HI-4460 Graphical Readout (Archived) User Manual
Page 26
![background image](/manuals/636338/26/background.png)
26
| Screens
Logging Screens
Use the up or down MODE key to select the first of four logging screens. The
common information is similar to that of the monitoring screens:
• A bar graph at the top shows the
field strength.
• The first line of text shows the
lower and upper limits of the
present range, as well as the peak
value.
• The next line shows the axis being
monitored and the selected scale.
• At the bottom is a selection of
menus that may be executed
using the four function keys.
The word LOG will appear when logging is active or scheduled. It appears
steady if conditions are not correct for a reading, such as not within the
specified levels or time frame, and it will flash on and off when it is recording
data. Alarm messages are also displayed in this area.
The first logging screen shows the selected log mode along with the logged
time, time remaining, and the log rate. The time remaining indicates the amount
of time before the memory fills up when in the continuous mode. When in
periodic or scheduled mode, it shows the time left for the current period of
logging.
Press the BACK/ENTER key or SCREEN/SELECT key to scroll through the
logging screens.
Archived 4/2/10