Section 3 | installation requirements, Requirements, Recommendations – Dako Autostainer User Manual
Page 15

Dako Autostainer
|
User Guide
15
Section 3
|
Installation Requirements
For installation of the Dako Autostainer the following requirements should be met to ensure that the
instrument functions properly.
Requirements
Surface
Prior to unpacking the Autostainer, ensure that the area to be used for
instrument is a solid, level surface that can safely support the weight of the
system (approximately 240 lbs. for instrument plus computer.)
Size
Minimum dimensions for the Autostainer work area are 2.00m wide x 0.8m
deep x 1.0m height (78” w x 30” d x 37” h) clearance.
Environment
The work area must also be an environment with an ambient temperature
between 18
°C-26°C (64°F-79°F), not facing direct sunlight and not
affected by a hood or vent.
Recommendations
Power source
Dako recommends a dedicated power source for the Autostainer to
prevent interference from other instruments or equipment.