Input power

100–240 VAC, 50/60 Hz

Fuse

2 ea. 2 Amp time delay 5 x 20 mm

Power input

IEC 60320 standard cord set with

ground

Power output

Voltage

0, 50–10,000 V, 1 V increments

Current

0–100 µA, 1 µA intervals

Power

0–1 W per lane

Peltier platform
Tray capacity

1 tray

Temperature

10–25ºC ±1.0ºC @ max ambient 23ºC

18–25ºC ±1.0ºC @ max ambient 31ºC

Focusing trays

Material Polycarbonate

IPG strip length

7, 11, 13, 17, 18, and 24 cm

Capacity

1–12 IPG strips per focusing tray

Channel volume

7 cm—7 ml, 11 cm—10 ml,

13 cm—11.2 ml, 17 cm—14.2 ml,

18 cm—15.2 ml, 24 cm—20.2 ml

Rehydration/equilibration trays

Material Polystyrene

Capacity

1–12 IPG strips per tray

IPG strip length

7, 11, 13, 17, 18, and 24 cm

Channel volume

7 cm—6.8 ml, 11 cm—9.6 ml,

13 cm—10.5 ml, 17 cm—14.2 ml,

18 cm—16 ml, 24 cm—19 ml

Environmental requirements
For indoor use only, at altitudes up to 6,000 ft
Operate at 10–35ºC ambient, with maximum
90% relative humidity
Regulatory

Safety

EN 61010-1:2001, IEC 61010-1:2001

Use NRTL to test for compliance to

UL61010-1:2004 and CAN/CSA

C22.2 No. 61010-1-04

EMC

EN61326 (1997 w/A1:98) Class A

FCC Code of Federal Regulations,

Title 47, Part 15, Subpart B, Class A

Other approvals

RoHS/WEEE Research Materials to

determine level of EFUP

Dimensions 7.3 in x 13.6 in x 15.1 in
Weight

19 lbs

User Interface

Display

QVGA resolution (320 x 240) touch

screen or mouse control

Programmable

Yes

Ramping

Step, linear, gradual, and hold voltage

ramping for each focusing step.

Hold mode as a final step to prevent

diffusion when IEF is complete

Protocol

2 GB

capacity

Data collection

.dat format