Appendix f specifications, Appendix f specifications 52 – Bio-Rad PROTEAN® i12™ IEF System User Manual
Page 56
PROTEAN i12 IEF System Instruction Manual
52 |
Appendix F
Specifications
Input power
100–240 VAC, 50/60 Hz
Fuse
2 ea. 2 Amp time delay 5 x 20 mm
Power input
IEC 60320 standard cord set with
ground
Power output
Voltage
0, 50–10,000 V, 1 V increments
Current
0–100 µA, 1 µA intervals
Power
0–1 W per lane
Peltier platform
Tray capacity
1 tray
Temperature
10–25ºC ±1.0ºC @ max ambient 23ºC
18–25ºC ±1.0ºC @ max ambient 31ºC
Focusing trays
Material Polycarbonate
IPG strip length
7, 11, 13, 17, 18, and 24 cm
Capacity
1–12 IPG strips per focusing tray
Channel volume
7 cm—7 ml, 11 cm—10 ml,
13 cm—11.2 ml, 17 cm—14.2 ml,
18 cm—15.2 ml, 24 cm—20.2 ml
Rehydration/equilibration trays
Material Polystyrene
Capacity
1–12 IPG strips per tray
IPG strip length
7, 11, 13, 17, 18, and 24 cm
Channel volume
7 cm—6.8 ml, 11 cm—9.6 ml,
13 cm—10.5 ml, 17 cm—14.2 ml,
18 cm—16 ml, 24 cm—19 ml
Environmental requirements
For indoor use only, at altitudes up to 6,000 ft
Operate at 10–35ºC ambient, with maximum
90% relative humidity
Regulatory
Safety
EN 61010-1:2001, IEC 61010-1:2001
Use NRTL to test for compliance to
UL61010-1:2004 and CAN/CSA
C22.2 No. 61010-1-04
EMC
EN61326 (1997 w/A1:98) Class A
FCC Code of Federal Regulations,
Title 47, Part 15, Subpart B, Class A
Other approvals
RoHS/WEEE Research Materials to
determine level of EFUP
Dimensions 7.3 in x 13.6 in x 15.1 in
Weight
19 lbs
User Interface
Display
QVGA resolution (320 x 240) touch
screen or mouse control
Programmable
Yes
Ramping
Step, linear, gradual, and hold voltage
ramping for each focusing step.
Hold mode as a final step to prevent
diffusion when IEF is complete
Protocol
2 GB
capacity
Data collection
.dat format