Bio-Rad PROTEAN® i12™ IEF System User Manual
Page 40
Chapter 7 Troubleshooting
36 |
PROTEAN i12 IEF System Instruction Manual
36 |
Problem
Cause
Solution
No current in a lane
Poor contact between IPG strip
and electrode
Make sure the gel side of the IPG
strip is in direct contact with the
electrode. For the gel-side down
configuration, make sure to
use the IPG strip retainers. See
Chapter 2 for proper placement
of IPG strips
No IPG strip in lane
Make sure that the lanes
selected for the run contain IPG
strips before starting a run
Incomplete wetting of electrode
wick
Wet the electrode wicks
completely as instructed in
Chapter 2
Incomplete rehydration of IPG
strip
Check the rehydration volumes
and times for the IPG strips used
No current in any lane
No contact between electrode
assembly and IPG strips
Make sure:
■
Electrode assembly is properly
seated in focusing tray
■
IPG strips are positioned
correctly, (for example, that
gel is in direct contact with the
electrode)
No contact between electrode
assembly and instrument
Make sure:
■
Gold contact pin of negative (-)
assembly is in direct contact
with cathode bar on instrument
■
Positive (+) assembly is
completely inserted into anode
of instrument
No IPG strips in lanes
Make sure IPG strips are
positioned in lanes before
starting a run
No conductivity in IPG strips
Make sure IPG strips are
rehydrated with correct reagents
Table 7.1. Troubleshooting guide.