Chapter 7 Troubleshooting

36 |

PROTEAN i12 IEF System Instruction Manual

36 |

Problem

Cause

Solution

No current in a lane

Poor contact between IPG strip
and electrode

Make sure the gel side of the IPG
strip is in direct contact with the
electrode. For the gel-side down
configuration, make sure to
use the IPG strip retainers. See
Chapter 2 for proper placement
of IPG strips

No IPG strip in lane

Make sure that the lanes
selected for the run contain IPG
strips before starting a run

Incomplete wetting of electrode
wick

Wet the electrode wicks
completely as instructed in
Chapter 2

Incomplete rehydration of IPG
strip

Check the rehydration volumes
and times for the IPG strips used

No current in any lane

No contact between electrode
assembly and IPG strips

Make sure:

Electrode assembly is properly
seated in focusing tray

IPG strips are positioned
correctly, (for example, that
gel is in direct contact with the
electrode)

No contact between electrode
assembly and instrument

Make sure:

Gold contact pin of negative (-)
assembly is in direct contact
with cathode bar on instrument

Positive (+) assembly is
completely inserted into anode
of instrument

No IPG strips in lanes

Make sure IPG strips are
positioned in lanes before
starting a run

No conductivity in IPG strips

Make sure IPG strips are
rehydrated with correct reagents

Table 7.1. Troubleshooting guide.