Chapter 3 Running a Protocol

18 |

PROTEAN i12 IEF System Instruction Manual

3.2 Screen Details

18 |

Screen

Details/Procedure

Main

Run — opens the protocol assignment screen for selection of
protocol(s) stored in the internal memory or on a USB flash drive
(only preprogrammed Bio-Rad protocols or saved .prt protocol files
appear).

Protocol Assignment Screens

Protocol Assignment

Display the available lanes (in a 12- or 6-lane format), available
memory devices, and Bio-Rad folder (with preprogrammed
protocols).

To assign a protocol to a lane(s):

1. Touch any lane to select it. Selected lanes are highlighted in blue.
2. Select the memory device/folder for display of stored protocols.
3. Select the protocol, which is highlighted and immediately assigned

to the highlighted lane(s). To select a preprogrammed protocol,
select the Bio-Rad folder and then select the protocols for the IPG
strip length. The protocols appear in the content box.

4. Repeat this procedure to assign different protocols to different

lanes.

Lanes Assigned

Displays the name of the assigned protocol in each selected
(highlighted) lane.

View Protocol — displays the details of the protocol highlighted in
the content box

Run — opens the Run Settings screen