beautypg.com

Bio-Rad PROTEAN® i12™ IEF System User Manual

Page 51

background image

Appendix C IEF Protocols

PROTEAN i12 IEF System Instruction Manual

| 47

| 47

Step

Voltage (V)

Ramp

Time

Units

1

8,000

Rapid

26,000

Volt Hr

2

750

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:20

HH:MM

2

8,000

Gradual

1:00

HH:MM

3

8,000

Rapid

26,000

Volt Hr

4

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

8,000

Rapid

20,000

Volt Hr

2

750

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:20

HH:MM

2

8,000

Gradual

1:00

HH:MM

3

8,000

Rapid

32,000

Volt Hr

4

750

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:20

HH:MM

2

8,000

Gradual

1:00

HH:MM

3

8,000

Rapid

40,000

Volt Hr

4

750

Hold

Step

Voltage (V)

Ramp

Time

Units

1

8,000

Rapid

29,000

Volt Hr

2

750

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:20

HH:MM

2

8,000

Gradual

1:00

HH:MM

3

8,000

Rapid

29,000

Volt Hr

4

750

Hold

11cm pH5.5–6.7
11cm pH6.3–8.3

11cm pH7–10 R

11cm pH7–10 G

11cm pH3–10 R
11cm pH3–10 NL R
11cm pH4–7 R
11cm pH5–8 R

11cm pH3–10 G
11cm pH3–10 NL G
11cm pH4–7 G
11cm pH5–8 G

11cm pH3–6 R

11cm pH3.9–5.1
11cm pH4.7–5.9

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:20

HH:MM

2

8,000

Gradual

1:00

HH:MM

3

8,000

Rapid

20,000

Volt Hr

4

750

Hold

11cm pH3–6 G