Chapter 5 setting defaults and managing folders, 1 setting default parameters (settings), R 5 setting defaults and managing folders 27 – Bio-Rad PROTEAN® i12™ IEF System User Manual
Page 31: Setting defaults and managing folders
PROTEAN i12 IEF System Instruction Manual
| 27
| 27
PROTEAN i12 IEF System Instruction Manual
Setting Defaults and
Managing Folders
5
5.1 Setting Default Parameters (Settings)
| 27
Use the Settings options to set default parameters and to customize screens:
■
System Time — sets the month, day, year, or time. Use the right and left arrow
buttons to make adjustments
■
Log Interval (min) — sets the data collection frequency (1–15 min; default
5 min). Select the field and use the up and down arrows to change the value
■
Default Run Name — name automatically associated with a run, and the run
data are stored in the Bio-Rad Data folder in the internal memory when the run
is complete. Stored run data files with names identical to new files being stored
in the same location are overwritten
■
Rehydration Step — sets global rehydration step defaults. If the rehydration
step is selected, all rehydration step options appear in the Run Settings screen.
If not selected, the details can be assigned in the Run Settings screen
■
Pause after Rehydrate — select to include a pause after rehydration
■
Default rehyd. time — sets the rehydration time (0–99 hr; default 12 hr)
■
Rehyd. temperature — sets the rehydration temperature (10–25ºC; default
20ºC)
■
Rehyd. voltage — sets the voltage for rehydration (0 V, 50–100 V; default 0 V)