Appendix C IEF Protocols

46 |

PROTEAN i12 IEF System Instruction Manual

46 |

Bio-Rad Protocols

(All protocols have a 50 μA current limit, though 100 μA is possible.)

Step

Voltage (V)

Ramp

Time

Units

1

4,000

Rapid

15,000

Volt Hr

2

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:15

HH:MM

2

4,000

Gradual

1:00

HH:MM

3

4,000

Rapid

15,000

Volt Hr

4

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

4,000

Rapid

10,000

Volt Hr

2

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:15

HH:MM

2

4,000

Gradual

1:00

HH:MM

3

4,000

Rapid

20,000

Volt Hr

4

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:15

HH:MM

2

4,000

Gradual

1:00

HH:MM

3

4,000

Rapid

25,000

Volt Hr

4

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

4,000

Rapid

16,000

Volt Hr

2

500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:15

HH:MM

2

4,000

Gradual

1:00

HH:MM

3

4,000

Rapid

16,000

Volt Hr

4

500

Hold

7cm pH5.5–6.7
7cm pH6.3–8.3

7cm pH7–10 R

7cm pH7–10 G

7cm pH3–10 R
7cm pH3–10 NL R
7cm pH4–7 R
7cm pH5–8 R

7cm pH3–10 G
7cm pH3–10 NL G
7cm pH4–7 G
7cm pH5–8 G

7cm pH3–6 R

7cm pH3.9–5.1
7cm pH4.7–5.9

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:15

HH:MM

2

4,000

Gradual

1:00

HH:MM

3

4,000

Rapid

20,000

Volt Hr

4

500

Hold

7cm pH3–6 G