Appendix C IEF Protocols

PROTEAN i12 IEF System Instruction Manual

| 49

Step

Voltage (V)

Ramp

Time

Units

1

10,000

Rapid

60,000

Volt Hr

2

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:30

HH:MM

2

10,000

Gradual

2:00

HH:MM

3

10,000

Rapid

60,000

Volt Hr

4

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

10,000

Rapid

44,000

Volt Hr

2

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:30

HH:MM

2

10,000

Gradual

2:00

HH:MM

3

10,000

Rapid

70,000

Volt Hr

4

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:30

HH:MM

2

10,000

Gradual

2:00

HH:MM

3

10,000

Rapid

88,000

Volt Hr

4

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

10,000

Rapid

63,000

Volt Hr

2

1,500

Hold

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:30

HH:MM

2

10,000

Gradual

2:00

HH:MM

3

10,000

Rapid

63,000

Volt Hr

4

1,500

Hold

24cm pH5.5–6.7
24cm pH6.3–8.3

24cm pH7–10 R

24cm pH7–10 G

24cm pH3–10 R
24cm pH3–10 NL R
24cm pH4–7 R
24cm pH5–8 R

24cm pH3–10 G
24cm pH3–10 NL G
24cm pH4–7 G
24cm pH5–8 G

24cm pH3–6 R

24cm pH3.9–5.1
24cm pH4.7–5.9

Step

Voltage (V)

Ramp

Time

Units

1

250

Rapid

0:30

HH:MM

2

10,000

Gradual

2:00

HH:MM

3

10,000

Rapid

44,000

Volt Hr

4

1,500

Hold

24cm pH3–6 G