Page 83

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Anion – MSM-HC – MCS – Prep 3

■■■■■■■■

73

Note

The guard column may only be installed after the initial start-up (see
Chapter 4.1, page 76)
of the instrument. Until then, use the
6.2744.040 coupling instead of the guard and separation column.

Connecting and rinsing the guard column

1 Connecting the guard column

Caution

When inserting the guard column, always ensure that it is inserted
correctly corresponding to the flow direction (if indicated).

Remove locking caps and/or stoppers from the guard column.

Fasten the lower end of the guard column with a 6.2744.070
short PEEK pressure screw to the column input capillary .

Fasten the connection capillary (3.4224.240) enclosed with the
guard column to the upper end of the guard column with a
6.2744.070 short PEEK pressure screw.
Guard columns are also available which can be screwed with the
upper end directly onto the separation column.

2 Rinsing the guard column

Place beaker under the outlet capillary of the guard column.

Start the high pressure pump and rinse the guard column approx.
5 minutes with eluent. Set the flow according to the correspond-
ing column information sheet.

Switch off the high pressure pump again.