Figure 14, Installing tubing weighting and aspiration filter, Figure 15 – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual
Page 44: Eluent aspiration tubing fully equipped
3.8 Eluent
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
850 Professional IC Anion – MSM-HC – MCS – Prep 3
4 Installing tubing weighting and aspiration filter
1
2
3
4
5
Figure 14
Installing tubing weighting and aspiration filter
1
Eluent aspiration tubing 6.1834.080
2
Eluent bottle attachment 6.1602.160
3
Tubing weighting
From accessory set 6.2744.210.
4
Clamping screw
From accessory set 6.2744.210.
5
Aspiration filter 6.2821.090
With filter holder from accessory set
6.2744.210.
■
Slide the tubing weighting (14-3) onto the eluent aspiration tub-
ing (14-1).
■
Slide the clamping screw (14-4) onto the eluent aspiration tubing
(14-1).
■
■
Screw together clamping screw (14-4) and filter holder (13-1).
The end of the tubing must still touch the base of the filter.
Figure 15
Eluent aspiration tubing fully equipped.
5 Mounting eluent aspiration tubing to the eluent bottle
■
Insert the eluent aspiration tubing into the eluent bottle (16-10).