beautypg.com

Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual

Page 128

background image

6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

118

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

Problem

Cause

Remedy

MCS – vacuum pump
defective.

Contact the Metrohm Service.

Background conduc-
tivity too high

MSM-HC – not connected.

Connecting the MSM-HC (see Chapter 3.18,
page 62)
.

MCS – not connected.

Connect the MCS.

Incorrect eluent.

Change eluent (see Chapter 5.4.2.3, page 82).

MSM-HC – regeneration or
rinsing solution flow prob-
lems.

Check the flow of the regeneration and rinsing
solution (see Chapter 3.18.2, page 62).

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean contact surfaces of the column chip
(with alcohol).

Column chip defective.

1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.

Individual peaks
greater than expec-
ted

Sample – Carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

MSM-HC – No (or
insufficient) pump-
ing of regeneration
or rinsing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure .

Peristaltic pump – filter
blocked .

Replace the filter .

MSM-HC – backpressure
too high.

Clean the MSM-HC (see Chapter 5.13.3.2,
page 108)
or replace parts (see Chapter
5.13.3.3, page 110)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing .

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.16.4, page 114)
.