5.13 High Capacity Metrohm Suppressor Module (MSM-HC)

■■■■■■■■■■■■■■■■■■■■■■

108

■■■■■■■■

850 Professional IC Anion – MSM-HC – MCS – Prep 3

3 Connecting the MSM-HC to the IC system

Reconnect the MSM-HC to the IC system. If the capacity problems
remain, the MSM-HC rotor A must be replaced (see Chapter
5.13.3.3, page 110)
.

5.13.3.2

Cleaning the MSM-HC

It may be necessary to clean the MSM-HC in the following cases:

Increased backpressure onto the connection tubings of the MSM-HC.

Blockage of the MSM-HC which cannot be eliminated (solutions can
no longer be pumped through the MSM-HC).

Jamming of the MSM-HC which cannot be eliminated (MSM-HC can
no longer be switched over).

1

2

3

4

5

Figure 55

MSM-HC – Components

1

Union nut

2

MSM-HC connecting piece 6.2835.010

3

MSM-HC rotor A 6.2842.000

4

MSM-HC housing

5

Slot in the MSM-HC housing

Cleaning the MSM-HC

Clean the MSM-HC as follows:

1 Disconnecting the MSM-HC from the IC system

Switch off the instrument.

Disconnect the MSM-HC from the separation column, peristaltic
pump and detector.

2 Dismantling the MSM-HC

Unscrew union nut (55-1) from the MSM-HC housing (55-4).