2 cleaning the msm‑hc, Figure 55, Msm-hc – components – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual
Page 118
5.13 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
850 Professional IC Anion – MSM-HC – MCS – Prep 3
3 Connecting the MSM-HC to the IC system
■
Reconnect the MSM-HC to the IC system. If the capacity problems
remain, the MSM-HC rotor A must be replaced (see Chapter
5.13.3.3, page 110).
5.13.3.2
Cleaning the MSM-HC
It may be necessary to clean the MSM-HC in the following cases:
■
Increased backpressure onto the connection tubings of the MSM-HC.
■
Blockage of the MSM-HC which cannot be eliminated (solutions can
no longer be pumped through the MSM-HC).
■
Jamming of the MSM-HC which cannot be eliminated (MSM-HC can
no longer be switched over).
1
2
3
4
5
Figure 55
MSM-HC – Components
1
Union nut
2
MSM-HC connecting piece 6.2835.010
3
MSM-HC rotor A 6.2842.000
4
MSM-HC housing
5
Slot in the MSM-HC housing
Cleaning the MSM-HC
Clean the MSM-HC as follows:
1 Disconnecting the MSM-HC from the IC system
■
Switch off the instrument.
■
Disconnect the MSM-HC from the separation column, peristaltic
pump and detector.
2 Dismantling the MSM-HC
■