14 metrohm co2 suppressor (mcs), 15 conductivity measuring system, 13 high capacity metrohm suppressor module – Metrohm 850 Professional IC MSM-HC MCS Prep 3 User Manual
Page 135: Msm-hc), 14 metrohm co, Suppressor (mcs)
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
850 Professional IC Anion – MSM-HC – MCS – Prep 3
■■■■■■■■
125
7.13
High Capacity Metrohm Suppressor Module
(MSM-HC)
Resistance to sol-
vents
No restriction
Switching dura-
tion
typ.100 ms
Operating pres-
sure
2.5 MPa (25 bar), valve function prevents damage at overpressure
7.14
Metrohm CO
2
suppressor (MCS)
Material
fluoropolymer
Resistance to sol-
vents
No restriction (apart from PFC)
Vacuum
Working range
Microprocessor-controlled / stabilized
Build-up time
after start
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1…1.0 mL
7.15
Conductivity measuring system
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm single range
Noise
< 0.1 nS at 1 µS/cm
Deviations from
the linearity
< 1 % for conductivity values greater than 1…16 µS/cm (typical for
analyses with sequential suppression)
Drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 0.2 nS/cm typical for sequential suppression